Test Code Billings Clinic: 9094 Mayo: HIVQN HIV-1 RNA Detection and Quantification, Plasma
Reporting Name
HIV-1 RNA Detect/Quant, PPerforming Laboratory
Mayo Clinic Laboratories in RochesterUseful For
Quantifying plasma HIV-1 RNA levels (viral load) in individuals living with HIV-1:
-Before initiating antiretroviral therapy to obtain baseline viral load
-Who may have developed HIV-1 drug resistance while on antiretroviral therapy
-Who may be noncompliant with antiretroviral therapy
Monitoring HIV-1 disease progression before or during antiretroviral drug therapy
Method Name
Real-Time Reverse Transcription Polymerase Chain Reaction (RT-PCR)
Shipping Instructions
Ship specimen frozen on dry ice. If shipment will be delayed for more than 24 hours, freeze plasma specimen at -20 to -80° C until shipment and transport on dry ice.
Specimen Required
Supplies: Sarstedt Aliquot Tube, 5 mL (T914)
Collection Container/Tube: Lavender top (EDTA)
Submission Container/Tube: Plastic vial
Specimen Volume: 1.5 mL
Collection Instructions:
1. Centrifuge blood collection tube and aliquot plasma into a plastic vial per manufacturer's instructions (eg, centrifuge and aliquot within 2 hours of collection for BD Vacutainer tubes).
2. Freeze aliquoted plasma for shipment.
Specimen Type
Plasma EDTASpecimen Minimum Volume
0.8 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Plasma EDTA | Frozen (preferred) | 84 days | |
Refrigerated | 6 days |
Reject Due To
Gross hemolysis | OK |
Gross lipemia | OK |
Gross icterus | OK |
Reference Values
Undetected
Interpretation
This assay has a plasma HIV-1 RNA quantification result range of 20 to 10,000,000 copies/mL (1.30-7.00 log copies/mL).
An "Undetected" result indicates that the assay was unable to detect HIV-1 RNA in the plasma specimen tested.
A result of "<20 copies/mL" indicates that HIV-1 RNA is detected, but the level present is less than the lower quantification limit of this assay. Due to the increased sensitivity of this assay, patients with previously low or undetectable HIV-1 viral load may show increased or detectable viral load with this assay. However, the clinical implications of a viral load below 20 copies/mL remain unclear. Possible causes of such a result include very low plasma HIV-1 viral load present (eg, in the range of 1-19 copies/mL), very early HIV-1 infection (ie, <3 weeks from time of infection), or absence of HIV-1 infection (ie, false-positive).
A result of ">10,000,000 copies/mL" with the result comment of "HIV-1 RNA level is >10,000,000 copies/mL (>7.00 log copies/mL). This assay cannot accurately quantify HIV-1 RNA above this level" indicates that HIV-1 RNA is detected, but the level present is above the upper quantification limit of this assay.
For monitoring a patient's response to antiretroviral therapy, the US Department of Health and Human Services Panel on Antiretroviral Guidelines for Adults and Adolescents defines virologic failure as a confirmed viral load above 200 copies/mL, which eliminates most cases of viremia resulting from isolated blips or assay variability. Confirmed viral load rebound (ie, >200 copies/mL) on 2 separate tests obtained at least 2 to 4 weeks apart should prompt a careful evaluation of patient's tolerance of current drug therapy, drug-drug interactions, and patient adherence.
Day(s) Performed
Monday through Saturday
Report Available
1 to 3 daysTest Classification
This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.CPT Code Information
87536
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
HIVQN | HIV-1 RNA Detect/Quant, P | 70241-5 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
113581 | HIV-1 RNA Detect/Quant, P | 70241-5 |
Forms
If not ordering electronically, complete, print, and send 1 of the following forms with the specimen:
-Microbiology Test Request (T244)
-General Request (T239)
Testing Algorithm
The following algorithms are available:
-HIV Prenatal Testing Algorithm, Including Follow-up of Reactive Rapid Serologic Test Results