Test Code DOCS 11-Deoxycorticosterone, Serum
Reporting Name
11-Deoxycorticosterone, SPerforming Laboratory
Mayo Clinic Laboratories in RochesterUseful For
Diagnosis of suspected 11-hydroxylase deficiency, including the differential diagnosis of 11-beta-hydroxylase 1 (CYP11B1) versus 11-beta-hydroxylase 2 (CYP11B2) deficiency
Diagnosis of glucocorticoid-responsive hyperaldosteronism
Evaluating congenital adrenal hyperplasia newborn screen-positive children, when elevations of 17-hydroxyprogesterone are only moderate, suggesting possible 11-hydroxylase deficiency
Method Name
Liquid Chromatography Tandem Mass Spectrometry (LC-MS/MS)
Specimen Required
Supplies: Sarstedt Aliquot Tube, 5 mL (T914)
Collection Container/Tube:
Preferred: Red top
Acceptable: Serum gel
Submission Container/Tube: Plastic vial
Specimen Volume: 0.5 mL
Collection Instructions:
1. Morning (8 a.m.) specimen is preferred.
2. Centrifuge and aliquot serum into a plastic vial.
Specimen Type
SerumSpecimen Minimum Volume
0.4 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Refrigerated (preferred) | 21 days | |
Frozen | 21 days | ||
Ambient | 7 days |
Reject Due To
Gross hemolysis | Reject |
Gross lipemia | OK |
Gross icterus | OK |
Reference Values
≤18 years: <30 ng/dL
>18 years: <10 ng/dL
Interpretation
In 11-beta-hydroxylase 1 (CYP11B1) deficiency, serum concentrations of cortisol will be low (usually <7 mcg/dL for a morning collection). 11-Deoxycortisol and 11-deoxycorticosterone are elevated, usually to at least 2 to 3 times (more typically 20-300 times) the upper limit of the normal reference range for a morning blood collection. Elevations in 11-deoxycortisol are usually relatively greater than those of 11-deoxycorticosterone, because of the presence of intact 11-beta-hydroxylase 2 (CYP11B2). For this reason, serum concentrations of all potent mineralocorticoids (corticosterone, 18-hydroxycorticosterone, and aldosterone) are typically increased above the normal reference range. Plasma renin activity is correspondingly low or completely suppressed. Caution needs to be exercised in interpreting the mineralocorticoid results in infants younger than 7 days; mineralocorticoid levels are often substantially elevated in healthy newborns in the first few hours of life and only decline to near-adult levels by week 1.
Mild cases of CYP11B1 deficiency might require corticotropin (previously adrenocorticotrophic hormone: ACTH)1-24 stimulation testing for definitive diagnosis. In affected individuals, the observed serum 11-deoxycortisol concentration 60 minutes after intravenous or intramuscular administration of 250 microgram of ACTH1-24 will usually exceed 20 ng/mL or demonstrate at least a 4-fold rise. Such increments are rarely, if ever, observed in unaffected individuals. The corresponding cortisol response will be blunted (<18 ng/mL peak).
In CYP11B2 deficiency, serum cortisol concentrations are usually normal, including a normal response to ACTH1-24. 11-Deoxycorticosterone will be elevated, often more profoundly than in CYP11B1 deficiency, while 11-deoxycortisol may or may not be significantly elevated. Serum corticosterone concentrations can be low, normal, or slightly elevated, while serum 18-hydroxycorticosterone and aldosterone concentrations will be low in the majority of cases. However, if the underlying genetic defect has selectively affected 18-hydroxylase activity, corticosterone concentrations will be substantially elevated. Conversely, if the deficit affects aldosterone synthase function primarily, 18-hydroxycorticosterone concentrations will be very high.
Expression of the CYP11B2 gene is normally regulated by renin and not ACTH. In glucocorticoid-responsive hyperaldosteronism, the ACTH-responsive promoter of CYP11B1 exerts aberrant control over CYP11B2 gene expression. Consequently, corticosterone, 18-hydroxycorticosterone, and aldosterone are significantly elevated in these patients and their levels follow a diurnal pattern, governed by the rhythm of ACTH secretion. In addition, the high levels of CYP11B2 lead to 18-hydroxylation of 11-deoxycortisol (an event that is ordinarily rare, as CYP11B1, which has much greater activity in 11-deoxycortisol conversion than CYP11B2, lacks 18-hydroxylation activity). Consequently, significant levels of 18-hydroxycortisol, which normally is only present in trace amounts, might be detected in these patients. Ultimate diagnostic confirmation comes from showing direct responsiveness of mineral corticoid production to ACTH1-24 injection. Normally, this has little if any effect on corticosterone, 18-hydroxycorticosterone, and aldosterone levels. This testing may then be further supplemented by showing that mineralocorticoid levels fall after administration of dexamethasone. Sex steroid levels are moderately to significantly elevated in CYP11B1 deficiency and much less, or minimally, pronounced in CYP11B2 deficiency. Sex steroid levels in glucocorticoid-responsive hyperaldosteronism are usually normal.
Most untreated patients with 21-hydroxylase deficiency have serum 17-hydroxyprogesterone concentrations well in excess of 1000 ng/dL. For the few patients with levels in the range of higher than 630 ng/dL (upper limit of reference range for newborns) to 2000 or 3000 ng/dL, it might be prudent to consider 11-hydroxylase deficiency as an alternative diagnosis. This is particularly true if serum androstenedione concentrations are also only mildly to modestly elevated, and if the phenotype is not salt wasting but either simple virilizing (female) or normal (female or male). 11-Hydroxylase deficiency, particularly if it affects CYP11B1, can be associated with modest elevations in serum 17-hydroxyprogesterone concentrations. In these cases, testing for CYP11B1 deficiency and CYB11B2 deficiency should be considered and interpreted as described above. Alternatively, measurement of 21-deoxycortisol might be useful. This minor pathway metabolite accumulates in CYP21A2 deficiency, as it requires 21-hydroxylaion to be converted to cortisol but is usually not elevated in CYP11B1 deficiency since its synthesis requires via 11-hydroxylation of 17-hydroxyprogesterone.
Day(s) Performed
Tuesday
Report Available
3 to 10 daysTest Classification
This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.CPT Code Information
82633
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
DOCS | 11-Deoxycorticosterone, S | 1656-8 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
46922 | 11-Deoxycorticosterone, S | 1656-8 |
Testing Algorithm
For more information see Steroid Pathways.