Test Code ESTS Estradiol, Rapid, Immunoassay, Serum
Reporting Name
Estradiol Rapid, Immunoassay, SPerforming Laboratory
Mayo Clinic Laboratories in RochesterUseful For
Rapid assessment of ovarian status, including follicle development, for assisted reproduction protocols (eg, in vitro fertilization)
Establishing time of ovulation and optimal time for conception
Method Name
Electrochemiluminescent Immunoassay (ECLIA)
Ordering Guidance
This assay is for reproductive assessment (eg, IVF, conception). For other clinical indications, order EEST / Estradiol, Serum.
The preferred method for measurement of low serum estradiol concentrations in children, males, and postmenopausal females is liquid chromatography-tandem mass spectrometry (LC-MS/MS), order EEST / Estradiol, Serum.
This test should not be used when monitoring estradiol concentrations in patients being treated with Fulvestrant due to possible cross-reactivity. In these patients, estradiol concentrations should be measured using mass spectrometry; order EEST / Estradiol, Serum.
Specimen Required
Patient Preparation: For 12 hours before specimen collection, patient should not take multivitamins or dietary supplements (eg, hair, skin, and nail supplements) containing biotin (vitamin B7).
Supplies: Sarstedt Aliquot Tube, 5 mL (T914)
Collection Container/Tube:
Preferred: Serum gel
Acceptable: Red top
Submission Container/Tube: Plastic vial
Specimen Volume: 0.5 mL
Collection Instructions: Within 2 hours of collection, centrifuge and aliquot serum into a plastic vial.
Specimen Type
SerumSpecimen Minimum Volume
0.4 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Frozen (preferred) | 180 days | |
Refrigerated | 5 days | ||
Ambient | 24 hours |
Reject Due To
Gross hemolysis | Reject |
Gross lipemia | OK |
Gross icterus | OK |
Reference Values
Males: 10-40 pg/mL
Females
Premenopausal: 15-350 pg/mL*
Postmenopausal: <10 pg/mL
*Estradiol concentrations vary widely throughout the menstrual cycle
Interpretation
Optimal time for conception is within 48 to 72 hours following the midcycle estradiol peak. Serial specimens must be drawn over several days to evaluate baseline and peak estradiol levels. Low baseline levels and a lack of rise, as well as persistent high levels without midcycle rise, are indicative of anovulatory cycles.
For determining the timing of initiation of ovarian stimulation in in vitro fertilization (IVF) studies, low levels before stimulation are critical, as higher values often are associated with poor stimulation cycles. Before final human chorionic gonadotropin (hCG) stimulation at mid-IVF cycle, estradiol concentrations above 2000 to 3000 pg/mL are considered by some IVF specialists to be indicative of an increased likelihood of ovarian hyperstimulation and it may be advisable to consider withholding further hCG stimulation.
Estradiol (E2) concentrations below 200 pg/mL following midcycle stimulation (hCG or follicle-stimulating hormone [FSH]) are associated with very low pregnancy success rates.
Estradiol concentrations change during the menstrual cycle, as follows:
-less than 50 pg/mL before midfollicular phase
-250 to 500 pg/mL midcycle peak as the follicle matures
-Abrupt decrease after ovulation
-125 pg/mL peak during the luteal phase
Estrogen replacement in reproductive-age women should aim to mimic natural estrogen levels as closely as possible. E2 levels should be within the reference range for premenopausal women and luteinizing hormone and FSH should be within the normal range.
Day(s) Performed
Monday through Sunday
Report Available
Same day/1 to 2 daysTest Classification
This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.CPT Code Information
82670
LOINC Code Information
Test ID | Test Order Name | Order LOINC Value |
---|---|---|
ESTS | Estradiol Rapid, Immunoassay, S | 2243-4 |
Result ID | Test Result Name | Result LOINC Value |
---|---|---|
ESTS | Estradiol Rapid, Immunoassay, S | 2243-4 |